2G 755與760超導磁力儀有何區別?

發布時間:2021-11-27 會員:

2G 755與760超導磁力儀有何區別?

 

一、樣品入口的尺寸和儀器長度

755 可用于離散樣品和 U-channel 樣品(處理長巖芯樣品或 8 個標準單樣)兩種,樣品入口 4.2cm, 主機長度78cm, 外徑 32cm。

760 也可以測量離散樣品和 U-Channel 樣品,同時還可以測量直徑 7cm 以內的長巖芯樣品,其入口尺寸為7.6cm。主機長 100cm,外徑 32cm。

由于 760 的樣品入口尺寸較大,其采集線圈直徑也相應地要大一些,對于標準的 2.5cm 散樣,760 的靈敏度僅為 755的一半,因此對于標準古地磁樣品,755 系統靈敏度是合適的選擇。

如果想要測量直徑大于 4cm 的樣品,則需要采購 760。

對于 7cm 以上的樣品,也可向廠家訂制更大入口尺寸的超導磁力儀。

二、退磁線圈以及脈沖磁化(IRM)螺線管可產生的磁場

隨著系統入口的增大,其能夠產生的峰值場強也顯著降低,具體如下:

4.2cm 入口(755) 橫向退磁場為 1800 奧斯特,IRM 磁場為 10000 奧斯特。

7.6cm 入口(760)橫向退磁場為 900 奧斯特,IRM 磁場為 5000 奧斯特。